• Total Organic Carbon Analyzer
  • Total Organic Carbon Analyzer

Total Organic Carbon Analyzer

No.DI1000
DI 1000 TOC Analyzer — Patented by Tailin, adopts UV oxidation and direct conductance method to detect TOC in water by calculating TC minus TIC. Measuring range 0.001–1.000 mg/L, precision ±5%, 4-min analysis. USP & CP compliant. Ideal for deionized water, purified water, WFI, and pharmaceutical/biochemistry applications. Online monitoring with 320×234 color screen, RS232 and printer interface.
$1500.00
  • Total Organic Carbon Analyzer

SPECIFICATION

DI1000 Total Organic Carbon TOC Analyzer for Pharmaceutical High Purity Water Online Monitoring

The DI 1000 TOC analyzer is a patented product independently developed by Tailin. It detects total organic carbon (TOC) in water samples through calculation based on total carbon (TC) and total inorganic carbon (TIC).

Working Principle

The DI1000 TOC analyzer uses a UV lamp to oxidize organic substances and convert them into carbon dioxide. The instrument adopts the direct conductance method to detect carbon dioxide concentration and output results. During testing, TC and TIC are measured separately. The TOC value is calculated by the formula: TOC = TC-TIC.

Features

  1. Low power consumption
  2. Simple and fast operation
  3. Specially designed for high-purity water testing and online monitoring
  4. Automatically stores test records for the last six months, supports daily record query and test result printing
  5. Fast testing speed and short pipeline cleaning time
  6. Compact, lightweight, low energy consumption and easy to move
  7. Equipped with buzzer high-limit alarm function
  8. Testing compliant with USP and CP requirements
  9. Large 320*234 color screen with customizable interface
  10. RS232 digital interface and micro-printer interface

Technical Parameters

表格

Item Specification
Power supply 220V
Mains frequency 50Hz
Rated output 100W
Dimension 44CM×18CM×26CM
Measuring limit 0.001mg/L
Measuring precision ±5%
Measuring range 0.001mg/L~1.000mg/L
Analysis time 4min
Response time within 15 min
Sample temperature 1-95℃
Sample flow rate 0.5ml/min
Relative humidity ≤85%
Repeatability error ±3%
Zero drift ±5%
Range drift ±5%
Linearity ±5%

Applications

Deionized pure water, pharmaceutical purified water, water for injection, pharmacy and biochemistry industries.